Recall Z-1707-2026

Medline Industries, LP · initiated February 16, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

THORACIC ROBOTS, DYNJ908777B

Codes / lots: UDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735

Reason for recall

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

FDA-determined cause: Under Investigation by firm

Action taken

Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.

Details

Recall number
Z-1707-2026
Initiated
February 16, 2026
Posted by FDA
April 1, 2026
Product code
OEZ – Cardiovascular Procedure Kit
Quantity
14,379 kits total
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1705-2026 Class IIAirway Exam Kit, DYKE1796Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified
Z-1706-2026 Class IIMedline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2.…Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified
Z-1708-2026 Class IIMedline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT…Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified
Z-1709-2026 Class IIMedline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G…Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified
Z-1710-2026 Class IIMedline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2.…Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified
Z-1711-2026 Class IIMEDLINE ANTERIOR HIP PACK DYNJ64672BKits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated… 02/16/2026Open, Classified