Recall Z-1709-2022

Cook Incorporated · initiated July 19, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Codes / lots: UDI-DI: (01)00827002075184 Lot Numbers: 14063361, 14087089, 14087092, 14117033, 14120926, 14120928, 14151216, 14151218, 14153159, 14158614, 14159695, 14159696

Reason for recall

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

FDA-determined cause: Under Investigation by firm

Action taken

Cook notified consignees by Urgent Medical Device Recall Letter on July 19, 2022. Letter states reason for recall, health risk and action to take: Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com.

Details

Recall number
Z-1709-2022
Initiated
July 19, 2022
Product code
DQX – Wire, Guide, Catheter
Quantity
1997 units
Distribution
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Recalling firm
Cook Incorporated, Bloomington, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1706-2022 Class IIRoadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN)RFPC-35-180 ORDER NUMBER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1707-2022 Class IIRoadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RFSPC-35-180 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1708-2022 Class IIRoadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1710-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPW-35-150 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1711-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-18-150…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1712-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-18-80 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1713-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-150…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1714-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-80 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1715-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-150…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1716-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWAS-35-80…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1717-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-150…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified
Z-1718-2022 Class IIRoadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWS-35-80 ORDER…Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration… 07/19/2022Open, Classified