Recall Z-1714-2026

Medline Industries, LP · initiated February 27, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B

Codes / lots: Medline SKU DYNJ42367B UDI/DI each 10193489211511, UDI/DI case 40193489211512, Lot Numbers: 24FBJ579, 24DBL828, 24DBE055, 24CBG967, 24ABN254, 23LBI540, 23KBF222, 23JBM761, 23JBA270, 23HBY791, 23HBD212, 23GBJ794, 23FBK893, 23EBQ313, 23DBC465.

Reason for recall

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

FDA-determined cause: Under Investigation by firm

Action taken

Medline Industries, LP issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 2/27/2026 via USPS and email. The notice explained the issue, risk associated with syringe disconnection and loose connection, and request customers remove and destroy all Namic RA Syringes. Over-labels will be provided for any affected kits on hand stating that the affected syringes must be removed and discarded from further use. The only exception is in rare circumstances where angiography is urgently required, no alternative device is available, and postponing the procedure would place the patient at significant risk. Required Action: Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-1714-2026
Initiated
February 27, 2026
Posted by FDA
April 7, 2026
Product code
OGJ – Nerve Block Tray
Quantity
3120 kits
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1712-2026 Class INAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1713-2026 Class INAMIC Angiographic Control Syringe, labeled as follows: 1.…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1715-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1716-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1717-2026 Class INAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1718-2026 Class INAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1719-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1720-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1721-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1722-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1723-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1724-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1725-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1726-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1727-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1728-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1729-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified
Z-1730-2026 Class IMedline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA)…Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe… 02/27/2026Open, Classified