Recall Z-1716-2008

Caire, Inc. · initiated October 23, 2007

Terminated

Product

Generation 4 electronic liquid gauges used on the Caire Liberator, Stroller, Sprint, and Hi-Flow Portable oxygen unit.

Codes / lots: Liberator Base units - CBB300634XXXX - CBB300738XXXX. Includes units produced from week 34-2006 through week 38-2007.

Reason for recall

Moisture ingress into the electronic liquid level indicator may cause inaccurate liquid oxygen level readings. The oxygen could run out without the patient knowledge.

FDA-determined cause: Device Design

Action taken

A Technical Service Bulletin was sent to consignees on/about October 23, 2007 to notify customers of the potential issue. Technical support and information for ordering replacement Generation 4 liquid gauges were offered. For technical concerns or if you have not received the notification from the firm, contact U.S. Technical Service at 1-800-482-2473 or European Technical Service at +44 (0) 797 1260077.

Details

Recall number
Z-1716-2008
Initiated
October 23, 2007
Posted by FDA
August 26, 2008
Terminated
December 17, 2009
Product code
BYJ – Unit, Liquid-Oxygen, Portable
Quantity
3470
Distribution
U.S. Nationwide and the following foreign countries: Argentina, Austria, Brazil, Canada, China, Chile, Central America/Caribbean, Colombia, Costa Rica, Europe, Austria, Belgium, Denmark, France, Germany, Greece, Hungary, Irish Republic, Italy, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom, Mexico, South America, Peru and Uruguay.
Recalling firm
Caire, Inc., Marietta, GA

Official FDA recall record