Recall Z-1723-2018

Datascope Corporation · initiated October 19, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BEQ-T 25501 3/8" Better Bladder Pack, Custom Tubing Kit, Catalog No. 709000370

Codes / lots: Lot 3000060037

Reason for recall

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

FDA-determined cause: Process control

Action taken

Affected consignees were notified by letter delivered via FedEx on October 19, 2017. After CTI expanded their recall in March 2018, Getinge accordingly alerted additionally affected consignees by letter delivered via FedEx on April 5, 2018. The recall letter advised customers of the issue and referenced the CTI recall notification, which was included as an attachment to Getinge's communication. End-users were instructed to examine their inventory to determine if they had any of the affected catalog/lot numbers. Any affected product is to be immediately quarantined and returned to Getinge by contacting Customer Service at 1(888) 627-8383 (option 2, then option 2 again), Monday through Friday, 8AM-6PM. In addition, customers were asked to complete a response form to acknowledge receipt of the notification, and return to Getinge. After communications with some end-users who were concerned about maintaining inventory of custom tubing kits, Getinge revised their recall strategy to include an option for replacement of the recalled Better Bladder device. Customers who wished to receive a replacement device will be provided with the device and a notice which must be affixed to the recalled kit. This will ensure that at time of use, it will be obvious to the user that the Better Bladder device inside the kit must be switched out with the replacement device prior to use. The removed recalled Better Bladder device must then be sent back to the recalling firm, Getinge. This information was communicated to customers via telephone call or visit on 4/20/18, and a summary of the communication was sent as an amended recall notification to customers who stated during the phone call that they still had recalled product on 4/21/18. Maquet/Getinge Customer support may be reached at 1(888) 627-8383 (press option 2, then option 2), Monday through Friday, 8AM-6PM.

Details

Recall number
Z-1723-2018
Initiated
October 19, 2017
Terminated
April 24, 2020
Product code
DTN – Reservoir, Blood, Cardiopulmonary Bypass
Quantity
10
Distribution
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
Recalling firm
Datascope Corporation, Fairfield, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1706-2018 Class IIBEQ-T 9410 Lurie Chldns Small 1/4, Custom Tubing Kit, Catalog No. 709000430The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1707-2018 Class IISmall Patient ECC Pack BEQ-TOP 22301, Catalog No. 701049513The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1708-2018 Class IIAdult ECC Pack BEQ-TOP 22300, Catalog No. 701049504The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1709-2018 Class IIBEQ-TOP 5210 ECC Set - Respiratory, Custom Tubing Kit, Catalog No. 701055604R01The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1710-2018 Class IIBEQ-TOP 33500 3/8" ECC PACK, Catalog No. 701053752The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1711-2018 Class IIBEQ-TOP 40700 HUNTINGTON, Catalog No. 701062899The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1712-2018 Class IIBEQ-TOP 22702 ECC W/HMOD 30000, Catalog No. 701051179The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1713-2018 Class IIBO-TOP 20705 SMALL PATIENT 1/4, Catalog No. 701053486The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1714-2018 Class IIBEQ-TOP 22701, Catalog No. 701050790The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1715-2018 Class IIBEQ-TOP 24202 ADULT ECC, Catalog No. 701054295The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1716-2018 Class IIBEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No.…The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1717-2018 Class IIBEQ-TOP 49101 Childrens Large ECC Pack, Custom Tubing Kit, Catalog No. 701067069R01The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1718-2018 Class IIBEQ-TOP 8504 1/4X1/4 PACK W/QP, Custom Tubing Kit, Catalog No. 701066949R02The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1719-2018 Class IIBEQ-TOP 38900 ECC PACK, Custom Tubing Kit, Catalog No. 701056368The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1720-2018 Class IIBEQ-TOP 50900 Custom ECC Pack-Neonatal 1/4, Custom Tubing Kit, Catalog No. 709000069R02The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1721-2018 Class IIBEQ-TOP 27610 ECC PACK 3/8, Custom Tubing Kit, Catalog No. 701065622The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1722-2018 Class IIBEQ-T 25502 1/4 Better Bladder , Custom Tubing Kit, Catalog No. 709000434The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1724-2018 Class IIBEQ-TOP 25500 ECC 3/8 w/QiD, Custom Tubing Kit, Catalog No. 701052461The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1725-2018 Class IINICU ECC Pack BEQ-TOP 39202, Custom Tubing Kit, Catalog No. 701067313R01The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1726-2018 Class IIBEQ-TOP 24500 ADULT ECC, Custom Tubing Kit, Catalog No. 701050253The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1727-2018 Class IIECC Small - Respiratory BEQ-TOP 5210, Custom Tubing Kit, Catalog No. 701055604The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1728-2018 Class IIBEQ-TOP 22301, Custom Tubing Kit, Catalog No. 701049513The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1729-2018 Class IIBEQ-TOP 22300, Custom Tubing Kit, Catalog No. 701049504The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1730-2018 Class IIBEQ-TOP 41201 ECC 3/8 PUMP PAC, Custom Tubing Kit, Catalog No. 701064867The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1731-2018 Class IIBO-TOP 13901 RAPID RESPONSE, Custom Tubing Kit, Catalog No. 701054874The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1732-2018 Class IIBO-TOP 20700 SMALL PATIENT 3/8, Custom Tubing Kit, Catalog No. 701050966The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated
Z-1733-2018 Class IIBEQ-TOP 50903 Custom ECC Pack-Pedriatric , Custom Tubing Kit, Catalog No. 709000078The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer.… 10/19/2017Terminated