Recall Z-1728-2022

Exactech, Inc. · initiated August 11, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F, REF 130-28-26 c. ACUMATCH GXL 0 DEGREE LINER 28MM SZ G, REF 130-28-27 d. ACUMATCH GXL 0 DEGREE LINER 28MM SZ H, REF 130-28-28 e. ACUMATCH GXL 0 DEGREE LINER 28MM SZ J, REF 130-28-29 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ G, REF 130-32-27 g. ACUMATCH GXL 0 DEGREE LINER 32MM SZ H, REF 130-32-28 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ J, REF 130-32-29 g. ACUMATCH GXL 0 DEGREE LINER 36MM SZ H, REF 130-36-28 h. ACUMATCH GXL 0 DEGREE LINER 36MM SZ J, REF 130-36-29 i. ACUMATCH GXL 0 DEGREE LINER 36MM SZ K, REF 130-36-30 j. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ E, REF 138-28-25 k. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ F, REF 138-28-26 l. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ G, REF 138-28-27 m. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ H, REF 138-28-28 n. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ J, REF 138-28-29 o. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ H, REF 138-36-28 p. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ J, REF 138-36-29 q. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ K, REF 138-36-30 r. ACUMATCH GXL 15 DEG +5 LAT LINER SZG 38MM, REF 138-36-27 s. ACUMATCH GXL 15DEG LINER 28MM SZ D, REF 132-28-24 t. ACUMATCH GXL 15DEG LINER 28MM SZ E, REF 132-28-25 u. ACUMATCH GXL 15DEG LINER 28MM SZ F, REF 132-28-26 v. ACUMATCH GXL 15DEG LINER 28MM SZ G, REF 132-28-27 w. ACUMATCH GXL 15DEG LINER 28MM SZ H, REF 132-28-28 x. ACUMATCH GXL 15DEG LINER 28MM SZ J, REF 132-28-29 y. ACUMATCH GXL 15DEG LINER 28MM SZ K, REF 132-28-30 z. ACUMATCH GXL 15DEG LINER 32MM SZ G, REF 132-32-27 aa. ACUMATCH GXL 15DEG LINER 32MM SZ H, REF 132-32-28 bb. ACUMATCH GXL 15DEG LINER 32MM SZ J, REF 132-32-29 cc. ACUMATCH GXL 15DEG LINER 32MM SZ K, REF 132-32-30 dd.ACUMATCH GXL 15DEG LINER 36MM SZ H, REF 132-36-28 ee. ACUMATCH GXL 15DEG LINER 36MM SZ J, REF 132-36-29 ff. ACUMATCH GXL 15DEG LINER 36MM SZ K, REF 132-36-30 gg. ACUMATCH GXL EXT COV LINER 28MM SZ E, REF 134-28-25 hh. ACUMATCH GXL EXT COV LINER 28MM SZ F, REF 134-28-26 ii. ACUMATCH GXL EXT COV LINER 28MM SZ G, REF 134-28-27 jj. ACUMATCH GXL EXT COV LINER 28MM SZ H, REF 134-28-28 kk. ACUMATCH GXL EXT COV LINER 28MM SZ J, REF 134-28-29 ll. ACUMATCH GXL EXT COV LINER 28MM SZ K, REF 134-28-30 mm. MCS GXL LINER 5/15 DEG 4448X28, REF 104-28-43 nn. MCS GXL LINER 5/15 DEG 5052X28, REF 104-28-44 oo. MCS GXL LINER 5/15 DEG 5458X28, REF 104-28-45 pp. MCS GXL LINER 5/15 DEG 5458X32, REF 104-32-45 qq. MCS GXL LINER 5/15 DEG 6064X32, REF 104-32-46 rr. MCS GXL LINER 5/15 DEG 6064X36, REF 104-36-46 ss. MCS GXL LINER 5/15 DEG 6670X36, REF 104-36-47

Codes / lots: a. REF 130-28-25, GUDID 10885862021984, Serial Numbers: 802593, 802599, 803717, 803737, 841318, 841321, 1572107, 1579369, 1580046, 1583363, 1613871, 1613872, 1613873, 1618998, 1638490, 1638492, 1638494, 1990100, 2282432, 2282433, 2282434; b. REF 130-28-26, GUDID 10885862021991, Serial Numbers: 801865, 801867, 801870, 801872, 801882, 801883, 801886, 801902, 803686, 803690, 803691, 803694, 803695, 803701, 848210, 848213, 848217, 848221, 848222, 848223, 848226, 848232, 848233, 848236, 1677105, 1677106, 1677108, 1677109, 1677110, 1677112, 1677113, 1677114, 1677115, 1677116, 1677118, 1677119, 1677121, 1677122, 1677123, 1677124, 1677126, 1677127, 1677128, 1677129, 1677130, 1677131, 2515330, 2787553, 3871934, 3871935, 3871943; c. REF 130-28-27, GUDID 10885862022004, Serial Numbers: 801542, 801544, 801546, 801547, 801554, 801571, 801572, 801578, 801580, 804033, 804034, 804035, 804036, 862139, 862144, 862146, 1681578, 1681579, 1681580, 1681581, 1681583, 1703643, 1727491, 1727492, 1727493, 1727495, 1727496, 1727497, 1727498, 1727499, 1727500, 1727501, 1727502, 1727503, 1727504, 1727505, 1727506, 1727507, 1727508; d. REF 130-28-28, GUDID 10885862022011, Serial Numbers: 801765; e. REF…

Reason for recall

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

FDA-determined cause: Process control

Action taken

The firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices. The Customers were instructed to take the following actions: CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory. EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com. Reach out to Customer Service at cust.svc@exac.com to request a Returned Goods (RG) Number for ALL affected product requested to be returned. WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL. If you have any questions, call (888) 912-0403.

Details

Recall number
Z-1728-2022
Initiated
August 11, 2022
Product code
LPH – Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Quantity
11143 DEVICES
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
Recalling firm
Exactech, Inc., Gainesville, FL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1725-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. 0 DEG…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1726-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1727-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1729-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1730-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1731-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. GXL…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1732-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1733-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ALL…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified
Z-1734-2022 Class IIGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a.…Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the… 08/11/2022Open, Classified