Recall Z-1756-2017

ConMed Corporation · initiated February 28, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 100 VOLTS AC, REF/Catalog Number 7-900-100, Rx Only. There is no problem with the Hyfrecator 2000 Electrosurgical Unit. -- ASSEMBLED IN MEXICO --- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502

Codes / lots: Serial Numbers: 1540139, 15280024, 15420101, 15470153, 15510179, 16010007, 1540144, 15280026, 15420102, 15470161, 15510182, 16010008, 15250002, 15290032, 15430106, 15490169, 15510186, 16010009, 15250004, 15290033, 15430107, 15490170, 15510189, 16010010, 15250005, 15300038, 15430110, 15490172, 15510190, 16430108, 15250006, 15350041, 15430111, 15500173, 15510192, 16430109, 15250007, 15350043, 15430113, 15500174, 16010001, 16430112, 15250008, 15350044, 15430114, 15500175, 16010002, 15260010, 15360054, 15430115, 15500176, 16010003, 15260012, 15360057, 15460140, 15500177, 16010004, 15260013, 15360059, 15460142, 15510178, 16010005, 15280023, 15360060, 15460146, 15510179, 16010006.

Reason for recall

For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.

FDA-determined cause: Mixed-up of materials/components

Action taken

ConMed sent an URGENT: MEDICAL DEVICE RECALL Notices (dated 2/28/2017)to all affected US customers by USPS First Class Mail. International letters are in the process of translation and are scheduled to be sent by March 15, 2017. Customers were instructed to review their inventory for any of the affected product and were asked to complete the business reply form and return it to ConMed Corporation, 525 French Road, Utica, New York 13502. Customers with questions or requests should contact the Field Action Support Team at 1-800-448-6506, fax to 315-624-3225, or email hyfr2000@conmed.com. For questions regarding this recall call 315-624-3533.

Details

Recall number
Z-1756-2017
Initiated
February 28, 2017
Product code
GEI – Electrosurgical, Cutting & Coagulation & Accessories
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
Recalling firm
ConMed Corporation, Utica, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1757-2017 Class IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in…For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the… 02/28/2017Terminated
Z-1758-2017 Class IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in…For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the… 02/28/2017Terminated
Z-1759-2017 Class IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in…For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the… 02/28/2017Terminated
Z-1760-2017 Class IIHYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single…For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the… 02/28/2017Terminated