Recall Z-1759-2010

Beckman Coulter Inc. · initiated January 21, 2010

Terminated

Product

COULTER PrepPlus 2, Part Number: 378600

Codes / lots: All Serial Numbers are affected.

Reason for recall

The recall was initiated after Beckman Coulter confirmed reports of potential dilution of reagent vials loaded onto the Preppies or PrepPlus 2 systems. Affected systems have a syringe pump with part number A46748 (see illustrations below) which cause the probe to prematurely dispense Isoflow prior to arriving at the wash station. This condition only occurs with worklist panels which include both

FDA-determined cause: Other

Action taken

A Product Corrective Action (PCA) letter with attached PCA Response Form was sent on the week of January 25, 2010 to all customers who purchased the pipetting and diluting system: COULTER PrepPlus, or COULTER PrepPlus 2. The letter provides the customers with an explanation of the problem identified and an instruction on proper placing antibody reagent when using a worklist panel which involves both blood specimens and antibody reagents. Customers were also instructed to complete and return the PCA Response Form in addition to sharing the information contained in the PCA with their laboratory staff. If consignees had any questions they were told to contact their local Beckman Coulter Representative. Questions can be directed to the company's Customer Service office at (800) 526-7694.

Details

Recall number
Z-1759-2010
Initiated
January 21, 2010
Posted by FDA
June 4, 2010
Terminated
April 4, 2012
Product code
JQW – Station, Pipetting And Diluting, For Clinical Use
Quantity
193 units in total
Distribution
Domestically and Canada.
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1758-2010 COULTER PrepPlus, Part Number: 286600The recall was initiated after Beckman Coulter confirmed reports of potential dilution of reagent vials… 01/21/2010Terminated