Recall Z-1766-2017

Merit Medical Systems, Inc. · initiated March 3, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.

Codes / lots: Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers: H1068587S1,H1073974, H1078179, H1078181, H1083629, H1083630; Catalog Number: SRB29AC, Lot Numbers: H1068591S1, H1078183, H1083631; Catalog Number: SRB29MED, Lot Numbers: H1068589S1, H1078184, H1078185, H1083633.

Reason for recall

Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.

FDA-determined cause: Process control

Action taken

Merit Medical sent an Urgent Product Recall Notice on March 3, 2017, via letter for consignees to immediately stop using or distributing the affected lots and return them to Merit. For any questions, please contact Merit Customer Service at (801) 208-4381. For further questions regarding this recall, please call (801) 253-1600.

Details

Recall number
Z-1766-2017
Initiated
March 3, 2017
Posted by FDA
April 10, 2017
Terminated
June 21, 2017
Product code
DXC – Clamp, Vascular
Quantity
1,940 units total
Distribution
Nationwide, Canadian and other foreign consignees. No VA/govt/military consignees.
Recalling firm
Merit Medical Systems, Inc., South Jordan, UT

Official FDA recall record