Recall Z-1774-2026

Molded Products Inc · initiated February 17, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

See Luer Cap Set, MPC-130, set, administration, intravascular

Codes / lots: UDI +B144MPC1300/$$529105203895, Lot Number 20389

Reason for recall

Molded Products, Inc. is voluntarily conducting a medical device recall of the MPC-130 See Luer Cap Set based on a confirmed complaint of the threaded sleeve not being engaged and becoming un-attached.

FDA-determined cause: Process control

Action taken

Molded Products, Inc. issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 02/17/2026 via USPS First Class Mail. The notice explained the issue and requested the following: Please take the following actions: 1. Please inspect your inventory to determine if you have any of the affected lot. Quarantine the affected product and initiate a return to Molded Products, Inc. for credit. 2. Share this recall notification will all users of the product to ensure they are also aware of this recall. 3. Complete the attached Medical Device Recall Customer Response Form and return to the contact noted below whether or not you have any of the impacted material so that Molded Products may acknowledge your receipt of this notification and process your return and credit. For complaints, adverse events, or questions, contact Teresa Wooster, twooster@moldedproducts.com; Ph# 800-435-8957 ext. 236 Monday Friday 8:00am-4:30pm (CT), Fax# 800-227-7935.

Details

Recall number
Z-1774-2026
Initiated
February 17, 2026
Posted by FDA
April 6, 2026
Product code
FPA – Set, Administration, Intravascular
Quantity
26900 units
Distribution
US Nationwide distribution in the states of TN, TX, NV, IL, FL.
Recalling firm
Molded Products Inc, Harlan, IA

Official FDA recall record