Recall Z-1778-2010

Ohio Medical Corporation · initiated May 10, 2010

Terminated

Product

NPWT Small Mobile Bag No Logo; an adjunct to the MoblVac Portable Wound Care Pump; Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, IL 60031; Model 769709.

Codes / lots: model 769709, all units

Reason for recall

Ohio Medical has received a report of an explosive event with a MoblVac Portable Woundcare Pump housing where parts were separated while in an enclosed nylon transport bag during use.

FDA-determined cause: Other

Action taken

Ohio Medical sent Urgent Medical Device Recall letters dated 5/7/10 to the direct accounts on 5/10/10, with follow-up telephone calls on 5/11/10. The accounts were informed of the one report of an explosion of the pump, and were requested to take the pumps out of service and return the pumps to Ohio Medical. The accounts were requested to complete and fax back to Ohio Medical the enclosed acknowledgement form indicating the number of pumps being returned. Any questions were directed to Anita Christ at 847-855-6270. Follow-up letters dated 5/13/10 were sent to the accounts on 5/14/10, providing instructions for the return of the pumps and bags. Shipping boxes will be sent to the accounts based on the number of pumps reported on the acknowledgment form.

Details

Recall number
Z-1778-2010
Initiated
May 10, 2010
Posted by FDA
June 7, 2010
Terminated
August 19, 2011
Product code
BTA – Pump, Portable, Aspiration (Manual Or Powered)
Quantity
218 bags
Distribution
Nationwide and internationally to Qatar and Oman
Recalling firm
Ohio Medical Corporation, Gurnee, IL