Recall Z-1781-2020

Degania Medical Devices Pvt. Ltd. · initiated March 17, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101080MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.

Codes / lots: Lot Numbers: S18012796

Reason for recall

Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

FDA-determined cause: Nonconforming Material/Component

Action taken

An Urgent Field Safety Notice dated 3/17/20 was sent to customers. Please immediately follow the Required Customer Actions set fo rth below. 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Please promptly return the completed Recall Response fo rm to Degania Silicone by e-mail: kkayam@qco.net. 4. The disposition of the recalled goods should be scrap. We would ask you to destruct the goods at your facil ity and send us the Scrap confirmation letter. Should you have any questions about this subject, please contact Degania Silicone: Degonia Silicone Ltd. Degania Bet, 1513000, ISRAEL Katia Kayam, Cust omer compla ints Officer Tel: +972-529532059 E-mail: kkayam@qco.net

Details

Recall number
Z-1781-2020
Initiated
March 17, 2020
Terminated
October 25, 2024
Product code
EZL – Catheter, Retention Type, Balloon
Quantity
700
Distribution
Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
Recalling firm
Degania Medical Devices Pvt. Ltd., Gurgaon, Imt Manesa

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1779-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102201100880MD - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1780-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102201101063MD - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1782-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102201101463MD - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1783-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102201101663MD - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1784-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102201101863MD - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1785-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102202101663DO - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1786-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102202101863DO - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1787-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102203101463DO - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1788-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102203101663DO - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated
Z-1789-2020 Class IITemperature Sensing 100% Silicone Foley Catheter, Part Number 102203101863DO - Product…Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the… 03/17/2020Terminated