Recall Z-1796-2008

Mentor Texas, Inc · initiated April 21, 2008

Terminated

Product

Psi-Tec Syringe Pump Infiltration Tubing, 10'; 10/box; Part Number PT-INF-T; distributed by Mentor Texas, LP, Irving, TX 75038.

Codes / lots: Lot #: IN4900, IN5906, IN5916, IN5922, IN5928, IN5934, IN5942, IN5950, IN5956, IN5970, IN5974, IN5984, IN6986, IN6990, IN6996, IN6998, IN6108, IN6114, IN6126, IN6142, IN6158, IN6168, IN7176, IN7196, IN7202, IN7240, IN7268 and IN8292.

Reason for recall

During packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices.

FDA-determined cause: Other

Action taken

A Customer Letter was issued to all consignees by the firm on 4/21/2008. Consignees were informed that a recent testing of these products discovered a potential defect in the packaging configuration, which could result in a loss of package integrity. Consignees were asked to review inventory and dispose of product and contact firm for credit.

Details

Recall number
Z-1796-2008
Initiated
April 21, 2008
Posted by FDA
September 5, 2008
Terminated
July 29, 2010
Product code
FPK – Tubing, Fluid Delivery
Quantity
1,151 boxes.
Distribution
Class II Recall - Worldwide Distribution ---- including USA, Puerto Rico and countries of Argentina, Austria, Bermuda, Brazil, Canada, China, Costa Rica, Cyprus, Denmark, England, Germany, Greece, Guatemala, India, Italy, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Panama, Paraguay, Portugal, Slovak Republic, Slovenia, Spain, South Africa, South Korea, Switzerland, Taiwan…
Recalling firm
Mentor Texas, Inc, Irving, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1790-2008 Psi-Tec Tubing, 12'; 10/box; Part Number PT-5558; distributed by Mentor Texas, LP…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1791-2008 Autofuse Tubing, DIS SI; Part Number AFTU-DS; distributed by Mentor Texas, LP, Irving…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1792-2008 Hi-Vac Tubing, 8'; 10/box; Part Number B-5558; distributed by Mentor Texas, LP, Irving…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1793-2008 Lamis Tubing Set, 9'; 10/box; Part Number LAM-TUBING; distributed by Mentor Texas, LP…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1794-2008 Y Tubing Set, 9'; 10/box; Part Number LAM-Y; distributed by Mentor Texas, LP, Irving, TX…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1795-2008 Irrigation Tubing, 25'; Part Number 5120-25; distributed by Mentor Texas, LP, Irving, TX…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1797-2008 Contour Genesis Infiltration Tubing, 13'; Part Number CG-INF-T; distributed by Mentor…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1798-2008 Contour Genesis Inf/IR Tubing; Part Number CG-UT; distributed by Mentor Texas, LP…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated
Z-1799-2008 Contour Genesis Aspiration Tubing, 12'; Part Number CG-PT-5558; distributed by Mentor…During packaging integrity validation, it was found that various product packages could fail, compromising… 04/21/2008Terminated