Recall Z-1803-2018

Bard Peripheral Vascular Inc · initiated April 16, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.014", 150 cm, REF SK15014. Intended to support a guidewire during access into the vasculature.

Codes / lots: UDI: (01)00801741067907(17)191207(10)VTBZ0193, Lot Number: VTBZ0193

Reason for recall

Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting.

FDA-determined cause: Process control

Action taken

The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emailed to BPV.CustomerSupportCenter@crbard.com or faxed to 1-800-994-6772.

Details

Recall number
Z-1803-2018
Initiated
April 16, 2018
Posted by FDA
May 11, 2018
Terminated
March 2, 2021
Product code
DQY – Catheter, Percutaneous
Quantity
130 units
Distribution
Nationwide Distribution
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1802-2018 Class IIBARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF…Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the… 04/16/2018Terminated