Recall Z-1806-2025

Smith & Nephew, Inc. · initiated March 18, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

Codes / lots: UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034

Reason for recall

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

FDA-determined cause: Process control

Action taken

On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to FieldActions@smith-nephew.com If you have any questions email the firm at the email address above.

Details

Recall number
Z-1806-2025
Initiated
March 18, 2025
Posted by FDA
May 19, 2025
Product code
HSB – Rod, Fixation, Intramedullary And Accessories
Quantity
75
Distribution
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
Recalling firm
Smith & Nephew, Inc., Memphis, TN

Official FDA recall record