Recall Z-1810-2025

MEDLINE INDUSTRIES, LP - Northfield · initiated April 22, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C

Codes / lots: UDI-DI 10888277858060 (EA); 40888277858061 (CASE) KIT LOTS 24CBJ925 24DBT405 24FBO318

Reason for recall

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

FDA-determined cause: Nonconforming Material/Component

Action taken

On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-1810-2025
Initiated
April 22, 2025
Posted by FDA
May 31, 2025
Product code
OJG – Neurological Tray
Quantity
1853 total
Distribution
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1809-2025 Class IMedline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU…Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on… 04/22/2025Open, Classified
Z-1811-2025 Class IMedline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU…Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on… 04/22/2025Open, Classified