Recall Z-1816-2016

Biomet, Inc. · initiated April 1, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Various trauma and sports medicine instruments and implants. cannula, surgical, general & plastic surgery and accessories

Codes / lots: Part #'s 900360 900364 900360H 900360HI 900363H 900364I Lot #'s 606770 343280 420090 742780 763750 742780 565910 343280

Reason for recall

Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

FDA-determined cause: Equipment maintenance

Action taken

Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.

Details

Recall number
Z-1816-2016
Initiated
April 1, 2016
Posted by FDA
May 26, 2016
Terminated
May 23, 2017
Product code
GEA – Cannula, Surgical, General & Plastic Surgery
Distribution
Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Recalling firm
Biomet, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1815-2016 Class IIVarious trauma and sports medicine instruments and implants. guide, surgical, instrumentInsufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1817-2016 Class IIVarious trauma and sports medicine instruments and implants. Plate, fixation, bone.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1818-2016 Class IIVarious trauma and sports medicine instruments and implants. Arthroscope.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1819-2016 Class IIVarious trauma and sports medicine instruments and implants. Rod, fixation…Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1820-2016 Class IIVarious trauma and sports medicine instruments and implants. Reamer.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1821-2016 Class IIVarious trauma and sports medicine instruments and implants. Bit, drill.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1822-2016 Class IIVarious trauma and sports medicine instruments and implants. Pin, fixation, smooth.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1823-2016 Class IIVarious trauma and sports medicine instruments and implants. Screw, fixation, bone.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1824-2016 Class IIVarious trauma and sports medicine instruments and implants. Passer.Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1825-2016 Class IIVarious trauma and sports medicine instruments and implants. Appliance, fixation…Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1826-2016 Class IIVarious trauma and sports medicine instruments and implants. Spinal interlaminal…Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1827-2016 Class IIVarious trauma and sports medicine instruments and implants. Orthopedic manual surgical…Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated
Z-1828-2016 Class IIVarious trauma and sports medicine instruments and implants. Fastener, fixation…Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile… 04/01/2016Terminated