Recall Z-1823-2024

Fisher & Paykel Healthcare, Ltd. · initiated April 2, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

Codes / lots: REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017

Reason for recall

Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

FDA-determined cause: Nonconforming Material/Component

Action taken

On 4/2/24, recall notices were sent to biomedical and respiratory department customers informing them to do the following: 1) Quarantine affected devices and work with firm to arrange collection and replacement. 2) Should a patient need to continue to use the device, the patient is instructed to contact their physician. Distributors are asked to facilitate device collection and replacement for their customers: a) If affected product has been further distributed create a list of affected customers for tracking purposes. b) Use the firm provided templates to create recall letters and response forms. Send the recall notice and response form to your customers to whom your firm has distributed affected devices. c) This recall notice should be shared with anyone who needs to be aware both within your organization and forwarded to any organization/ where potentially affected devices have been transferred. 3) Complete and return the response form via email at FieldAction.USA@fphcare.com Firm is removing and replacing affected devices. If you have any questions, please contact firm Customer Care via email at FieldAction.USA@fphcare.com or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.

Details

Recall number
Z-1823-2024
Initiated
April 2, 2024
Posted by FDA
May 15, 2024
Product code
BTT – Humidifier, Respiratory Gas, (Direct Patient Interface)
Quantity
7,147
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Recalling firm
Fisher & Paykel Healthcare, Ltd., Auckland, East Tamak