Recall Z-1825-2023

Remote Diagnostic Technologies Ltd. · initiated May 8, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020

Codes / lots: UDI-DI: 07613365002737. Serial numbers prior to 7022.001634

Reason for recall

Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".

FDA-determined cause: Device Design

Action taken

On 5/8/23, Philips mailed correction notices to customers who were asked to do the following: Potential risks associated with partial or complete failure of the device, should a patient be involved, will be mitigated by immediately starting CPR/ chest compressions in case of device dysfunction/failure. Philips will contact to arrange for the device to be returned for rework. A loner device will be provided during the time of the rework. This notice needs to be shared with those who need to be aware within your organization and any organization where the potentially affected devices have been transferred. Complete and return the Customer/Distributor Reply Form: https://philips.efmfeedback.com/se/705E3ED8496AFDF8/ https://philips.efmfeedback.com/se/705E3ED8496AFDF8 Customers/Distributors who need further information or support can contact RDT.Recall.Response@Philips.com, or 1 800 722 9377 In addition, distributors were asked to forward the correction notice to their customers. The notice should be modified to include the distributor's contact information.

Details

Recall number
Z-1825-2023
Initiated
May 8, 2023
Posted by FDA
June 12, 2023
Product code
LDD – Dc-Defibrillator, Low-Energy, (Including Paddles)
Quantity
1147
Distribution
US Nationwide distribution in the states of UT, CO, CA, NE, IL, TX, WV, KS, GA.
Recalling firm
Remote Diagnostic Technologies Ltd., Farnborough