Recall Z-1831-2019

Covidien LLC · initiated May 17, 2019

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

EGIA45 CT AR MD THK REL, EGIA45CTAMT

Codes / lots: N8L0401KY

Reason for recall

The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.

FDA-determined cause: Process control

Action taken

All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item codes and lots. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology. " Complete the Recalled Product Return Form. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.

Details

Recall number
Z-1831-2019
Initiated
May 17, 2019
Product code
GDW – Staple, Implantable
Quantity
3,113,280
Distribution
Nationwide domestic distribution, worldwide foreign distribution.
Recalling firm
Covidien LLC, North Haven, CT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1825-2019 Class IIEGIA 30 ARTICULATING MED THICK SULU, Item Code EGIA30AMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1826-2019 Class IIEGIA 45 ARTICULATING MED THICK SULU, Item Code EGIA45AMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1827-2019 Class IIEGIA 45 ARTICULATING XTRA THICK SULU, Item Code EGIA45AXTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1828-2019 Class IIEGIA 60 ARTICULATING EXTRA THICK SULU, Item code EGIA60AXTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1829-2019 Class IIEGIA 60 ARTICULATING MED THICK SULU , EGIA60AMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1830-2019 Class IIEGIA 60 CURVED TIP ART MED THICK SULU, EGIA60CTAMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1832-2019 Class IISIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1833-2019 Class IISIGNIA TRI-STAPLE 2.0 BLACK 45MM EXTRA THICK RELOA, SIG45AXTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1834-2019 Class IISIGNIA TRI-STAPLE 2.0 BLACK 60MM EXTRA THICK RELOA, SIG60AXTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1835-2019 Class IISIGNIA TRI-STAPLE 2.0 CURVED TIP 45MM MEDIUM/ THIC , SIG45CTAMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified
Z-1836-2019 Class IISIGNIA TRI-STAPLE 2.0 CURVED TIP 60MM MEDIUM/ THIC , SIG60CTAMTThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in… 05/17/2019Open, Classified