Recall Z-1850-2008

Atrium Medical Corporation · initiated May 12, 2008

Terminated

Product

HydraGlide Heparin Coated Silicone Thoracic Catheters - Mediastinal, Eyes on Top & Bottom, Atrium Code # 14509, Size 9mm, # Eyelets 9, 10 per case

Codes / lots: Lot 10311260

Reason for recall

Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated chondroitin sulfate

FDA-determined cause: Nonconforming Material/Component

Action taken

Atrium Medical issued a press release on 5/12/2008 and notified consignees by facsimile on 5/12/2008. Customers were instructed to contact the firm for a return goods authorization and to notify the firm upon receipt of the recall notice.

Details

Recall number
Z-1850-2008
Initiated
May 12, 2008
Posted by FDA
August 14, 2008
Terminated
August 20, 2009
Product code
GBZ – Catheter, Cholangiography
Quantity
6 cases (10 individual/case)
Distribution
Worldwide Distribution: USA, US Virgin Islands, South Africa, Austria, Taiwan, Venezuela, and Saudi Arabia.
Recalling firm
Atrium Medical Corporation, Hudson, NH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1830-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1831-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1832-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1833-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1834-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1835-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1836-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1837-2008 HydraGlide Heparin coated PVC Straight Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1838-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparinc allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1839-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1840-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1841-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1842-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1843-2008 HydraGlide Heparin Coated PVC Right Angle Thoracic Catheters with Tapered Connector Tip…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1844-2008 ATP HydraGlide Trocar - Includes Tapered ATP Stylet, Atrium Code # 9408, Size (Fr) 8, #…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1845-2008 ATP HydraGlide Trocar - Includes Tapered ATP Stylet, Atrium Code # 9420, Size (Fr) 20, #…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1846-2008 ATP HydraGlide Trocar - Includes Tapered ATP Stylet, Atrium Code # 9428, Size (Fr) 28, #…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1847-2008 HydraGlide Heparin Coated Silicone Thoracic Catheters - Straight, Atrium Code # 14120…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1848-2008 HydraGlide Heparin Coated Silicone Thoracic Catheters - Straight, Atrium Code # 14124…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1849-2008 HydraGlide Heparin Coated Silicone Thoracic Catheters - Straight, Atrium Code # 14132…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1851-2008 HydraGlide Heparin Coated Silicone Thoracic Catheters - Mediastinal, Eyes on Top, Atrium…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly conminated with oversulfated… 05/12/2008Terminated
Z-1852-2008 HydraGlide XL Heparin Coated Silicone Thoracic Catheters, Increased number of eyelets…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated
Z-1853-2008 HydraGlide XL Heparin Coated Silicone Thoracic Catheters, Increased number of eyelets…Heparin-coated Thoracic Catheters were manufactured with heparin allegedly contaminated with oversulfated… 05/12/2008Terminated