Recall Z-1864-2011

Abbott Point Of Care Inc. · initiated December 17, 2010

Terminated

Product

Cardiac Marker Control Calibration Verification Set, List number 136604/06F15-02. Used to verify the accuracy of results over the measurement range of the i-STAT cTnI test.

Codes / lots: Cardiac Marker Control Calibration Verification Set - Lot number M10208

Reason for recall

Presence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set.

FDA-determined cause: Device Design

Action taken

Urgent Recall Notice letters with business reply cards were sent Fed ex on December 17, 2010 to all accounts. International notifications were sent on December 21, 2010. The letter identified the affected product and lot numbers along with a background explanation of the issue. The firm recommends that customers remove the cap from the vial; squeeze the vial 10 times, prior to thawing; replace cap on the vial; and then follow the remaining steps for use as outlines in section 14-12 of the i-STAT 1 System Manual. Customers are to complete and return the Business Reply card to acknowledge receipt of the letter. If the affected product had been further distributed then a copy of the letter should be provided to those customers as well. If additional information is needed, customers should contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or their Abbott Point of Care representative.

Details

Recall number
Z-1864-2011
Initiated
December 17, 2010
Posted by FDA
April 1, 2011
Terminated
October 16, 2012
Product code
JJX – Single (Specified) Analyte Controls (Assayed And Unassayed)
Quantity
557 units total
Distribution
Worldwide Distribution
Recalling firm
Abbott Point Of Care Inc., Princeton, NJ

Official FDA recall record