Recall Z-1864-2012

Hill-Rom, Inc. · initiated June 18, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Advanta 2 Bed. The Advanta 2 Bed is intended for low to moderate acuity patients in the medical/surgical area of the hospital.

Codes / lots: AdvantaTM 2 beds (P1190) produced between August 10, 2009 and October 6, 2011; bed S/N between K222AT9180 and M279AT4403

Reason for recall

During an evaluation of Advanta 2 siderail samples returned from the field it was found that the screws used to attach the siderail were being stripped from the plastic cavity of the siderail.

FDA-determined cause: Process design

Action taken

Hill-Rom sent an "URGENT FIELD SAFETY NOTICE" dated May 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included a Customer Response Form for customers to complete and return to the firm. Contact Hill-Rom Technical Support at 800-445-3720 for questions regarding this notice.

Details

Recall number
Z-1864-2012
Initiated
June 18, 2012
Posted by FDA
June 22, 2012
Terminated
June 6, 2014
Product code
FNL – Bed, Ac-Powered Adjustable Hospital
Quantity
7843 devices
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, France, and Singapore.
Recalling firm
Hill-Rom, Inc., Batesville, IN

Official FDA recall record