Recall Z-1872-2008

Biomet Microfixation, Inc. · initiated March 19, 2008

Terminated

Product

Lactosorb Trauma Plating System, part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. The system is used in trauma and reconstructive surgical procedures in the mid face and craniofacial skeleton. Used to support non-load bearing tissues of cranial maxillofacial anatomy, including but not limited to brow fixation.

Codes / lots: Part numbers: 915-2100, 915-2423, 915-2113, 915-2810, 915-2418, 915-2105, 915-2832 and CP-1323. Lot numbers: 898240, 898330, 898250, 898350, 898320, 951140, 868920, and 978810.

Reason for recall

The outer foil pouch on the Lactosorb Plate system was not properly sealed. The product is labeled as sterile, but the lot was not sterilized. These devices are intended to be implanted into bone and are required to be sterile. A heavy bioburden would increase the potential for bone infection (osteomyelitis).

FDA-determined cause: Packaging process control

Action taken

On 3/19/08 the firm issued an Urgent Medical Device Recall Notice to its consignees by certified letters with return receipt. All returned products will be held in quarantine until final disposition is determined. If you have any questions contact Rachel Osbeck at 1-800-874-7711 or (904) 741-4400 ext 448.

Details

Recall number
Z-1872-2008
Initiated
March 19, 2008
Posted by FDA
September 16, 2008
Terminated
September 11, 2009
Product code
JEY – Plate, Bone
Quantity
117
Distribution
Class II Recall - Worldwide Distribution --- USA including states of NC, MI, TX, GA, HI, OK, IN and CA, and countries of Denmark, Germany, Mexico, Italy, Brazil, Japan, Canada, South Africa, China and the Philippines.
Recalling firm
Biomet Microfixation, Inc., Jacksonville, FL