Recall Z-1887-2018
Ethicon Endo-Surgery Inc · initiated March 26, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm, 150mm shaft length, 2B5XT
Codes / lots: Lots P4TD1E R4001U R4052P
Reason for recall
Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.
FDA-determined cause: Process control
Action taken
The firm notified customers and distributors by letter on March 26, 2018, and March 27, 2018, respectively. Customers were informed of the issue and potential risk of prolonged surgery or introduction of minute particles. Customers were instructed to: 1. Examine your inventory immediately to determine if you have any products subject to this recall on hand and quarantine such product(s). 2. Remove the products subject to this recall from your inventory and communicate the issue to all relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any of the devices subject to this recall have been forwarded to another facility, please contact that facility to arrange return. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or Ethicon3546@stericycle.com within three (3) business days. Please return the BRF even if you do not have the product lot subject to this recall. 4. Customers are required to return all unused devices subject to this recall that are in their inventory immediately. 5. To return unused devices subject to this recall photocopy the completed BRF, place it in the box with the subject product(s), and affix the pre-paid authorized shipping label included with this recall notification letter. Distributors were instructed to: 1. Examine your inventory immediately to determine if you have the product lot subject to this recall on hand and quarantine such product(s). 2. Remove the product lot subject to this recall and communicate the issue to all relevant materials management personnel, or anyone else in your facility who needs to be informed. 3. Complete the Business Reply Form confirming receipt of this notice and fax or email it to Stericycle at 1-888-965-5794 or Ethicon3546@stericycle.com within three (3) business days. Please return the BRF even if you do not have the product lot subject to this rec
Details
- Recall number
- Z-1887-2018
- Initiated
- March 26, 2018
- Terminated
- April 8, 2020
- Product code
- GCJ – Laparoscope, General & Plastic Surgery
- Quantity
- 1500
- Distribution
- Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.
- Recalling firm
- Ethicon Endo-Surgery Inc, Blue Ash, OH
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1885-2018 Class II | 5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm…Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition… | 03/26/2018Terminated |
| Z-1886-2018 Class II | 5mm Endopath Xcel with Optiview Technology, Bladeless Trocar with Stability Sleeve, 5mm…Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition… | 03/26/2018Terminated |
| Z-1888-2018 Class II | 5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability Sleeve, 5mm…Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition… | 03/26/2018Terminated |
| Z-1889-2018 Class II | 5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability Sleeve, 5mm…Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition… | 03/26/2018Terminated |