Recall Z-1891-2017

Perouse Medical · initiated April 5, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

DOLPHIN INFLATION DEVICE CALIBER INFLATION DEVICE; Catalog Number: CL3030- 018SNA- 018SND 0185NA 0185ND; INTERNAL CODE NUMBER: 0185TR- 0185NA- 018SND 0185NA 0185ND

Codes / lots: 1 5045606 - 15055430 - 15065340 - 15065685 -15065744- 15065858- 15065970- 15075121 - 15075265 - 15075411- 15075821 - 1 5075851 - 15075942 - 15085001 - 15085075 -15085378- 15085435 -15085515- 15085555 -15105052- 15125214- 16015267- 16025034- 16025640

Reason for recall

Complaints regarding broken blister.

FDA-determined cause: Packaging change control

Action taken

Consignees were notified of the recall on April 5, 2017. They were informed of the issue and asked to take the following steps: 1. Inspect your stock to determine if you still have in your ownership devices among which the references and batch codes are listed in front page; 2. Control the potential damage of the blister according to the pictures above and the instructions mentioned on the labeling and the Instruction For Use: any damaged and/or cracked blister must be immediately discarded; 3. Use the compliant products without risk, after control. Indeed, the primary packaging will not deteriorate in time. If, to date, there is no crack in the blister, then the integrity of the product is not compromised; 4- Do not use the non-compliant products; 5. Complete and sign the attached reply form, then return it within 5 working days following the reception of this letter, and keep a copy; 6. Contact customer service for the modalities of return, replacement and associated support at +33 (0)4.72.39.74.13 or by fax at +33 (0}3.44.08.17.67 or by email at iblayon@vvgon.com (CC mhpourriere@vygon.com); 7. Communicate the safety information to any person concerned in your company and to all end customers to which the product were transferred.

Details

Recall number
Z-1891-2017
Initiated
April 5, 2017
Terminated
February 1, 2021
Product code
MAV – Syringe, Balloon Inflation
Quantity
18,431
Distribution
US: AZ (shipped through Belgium to USA), MN, TN
Recalling firm
Perouse Medical, Ivry Le Temple

Official FDA recall record