Recall Z-1891-2026
Medline Industries, LP · initiated February 27, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes ARTERIOGRAM TRAY 00-402001T, 00-402001U CATH LAB-VEIN PROCEDURES DYNJ60329A CENTRAL VENOUS ACCESS PACK-LF CVI5070 FISTULAGRAM PACK DYNJ66259 INVASIVE LINE INSERTION DYNJ44123C NON VASCULAR PACK DYNJ68037B, DYNJ68037C PICC ABSCESS PACK-LF DYNJ0774005G PICC LINEPACK (PCLUI)642-LF DYNJ47717B PICC PHC DYNJ37484C
Codes / lots: 00-402001T UDI-DI 10193489939682 (EA) 40193489939683 (CS) LOTS 23EBT729 23GBR995 23GBS336 00-402001U UDI-DI 10195327517380 (EA) 40195327517381 (CS) LOTS 23JBU819 23LBG449 24ABS585 24CBF190 24DBA960 24EBB048 24FBN231 24HBA892 24IBD316 24IBS824 DYNJ60329A UDI-DI 10193489972221 (EA) 40193489972222 (CS) LOT 24ABP714 CVI5070 UDI-DI 10195327526061 (EA) 40195327526062 (CS) LOT 24FBF359 DYNJ66259 UDI-DI 10193489325485 (EA) 40193489325486 (CS) LOT 22FMB867 DYNJ44123C UDI-DI 10195327599140 (EA) 40195327599141 (CS) LOTS 24ABQ097 24DMH637 24FMG692 24GMA523 24HME792 24JMA878 24JMI393 DYNJ68037B UDI-DI 10195327476458 (EA) 40195327476459 (CS) LOT 24HBK116 DYNJ68037C UDI-DI 10198459239328 (EA) 40198459239329 (CS) LOT 25BBO522 25DBR880 DYNJ0774005G UDI-DI 10193489271485 (EA) 40193489271486 (CS) LOTS 23DMC237 23EMC431 23FMG987 23HBL979 23JBG840 23KBJ401 23LBR858 24CBB784 24EBA777 DYNJ47717B UDI-DI 10195327355302 (EA) 40195327355303 (CS) LOTS 23GMB288 23HMA341 23HMH322 DYNJ37484C UDI-DI 10195327397340 (EA) 40195327397341 (CS) LOTS 24EDB695 24HDC050 24IMH716 24JMH721 24KMG262 ***Updated 5/20/26 - The following lots were included in the customer notification but not the FDA submission.*** 00-402001T…
Reason for recall
Unapproved design changes to the products outside of the 510(k) clearance.
FDA-determined cause: Component change control
Action taken
On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Details
- Recall number
- Z-1891-2026
- Initiated
- February 27, 2026
- Posted by FDA
- April 17, 2026
- Product code
- OFF – Central Venous Catheter Tray
- Quantity
- 270,311 total
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
- Recalling firm
- Medline Industries, LP, Northfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1876-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1877-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1878-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1879-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1880-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1881-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1882-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1883-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1884-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1885-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1886-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1887-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1888-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1889-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes E…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1890-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |
| Z-1892-2026 Class II | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes…Unapproved design changes to the products outside of the 510(k) clearance. | 02/27/2026Open, Classified |