Recall Z-1901-2025

Beckman Coulter, Inc. · initiated May 5, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Access Testosterone assay, Catalog Number 33560

Codes / lots: UDI-DI: 15099590231347; Serial Number: 538081

Reason for recall

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

FDA-determined cause: Equipment maintenance

Action taken

An URGENT MEDICAL DEVICE RECALL notification letter dated 4/23/25 was sent to customers. ACTION: Discontinue using and discard all remaining Access Testosterone reagent packs from the lot listed above. Beckman Coulter recommends performing a retrospective review on patient results generated with the Access Testosterone lot listed above and assessing the need for retesting. Contact your local Beckman Coulter representative for replacement product requests. Beckman Coulter is no longer distributing the affected Access Testosterone reagent pack lot number and has corrected the manufacturing issue that led to cross-contamination. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com If you need replacement product: o Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821

Details

Recall number
Z-1901-2025
Initiated
May 5, 2025
Posted by FDA
June 2, 2025
Product code
CDZ – Radioimmunoassay, Testosterones And Dihydrotestosterone
Quantity
70 units
Distribution
US State: FL. UAE, Vietnam
Recalling firm
Beckman Coulter, Inc., Chaska, MN

Official FDA recall record