Recall Z-1902-2026
Olympus Corporation of the Americas · initiated March 25, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2114.
Codes / lots: Model Number: MAJ-2114. UDI Number: 04953170404092. Lot Numbers: 09A, 0XA, 0YA, 0ZA, 11A, 12A, 24A, 25A, 26A, 27A, 28A, 29A, 2XA, 2YA, 2ZA, 31A, 32A, 33A, 34A, 35A, 36A, 37A, 38A, 39A, 3XA, 3YA, 3ZA, 41A, 42A, 43A, 44A, 45A, 46A, 47A, 48A, 49A, 4XA, 4YA, 4ZA, 51A.
Reason for recall
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
FDA-determined cause: Device Design
Action taken
Olympus notified consignees on about 03/25/2026 via letter titled "URGENT: MEDICAL DEVICE REMOVAL." Consignees were notified of the issue and hazard involved and were instructed to examine inventory and OER-ELITE devices for the affected product. The current version has black lock levers. The previous versions have gray lock levers or integrated lock levers, which are the same color as the rest of the connector. Consignees were instructed that if they have sufficient quantities of the current version (black) to support reprocessing needs, stop using the previous version and set aside for return. They may proceed with using the current version, only. If they have insufficient quantities of the current version (black) available, and do not have another method of endoscope reprocessing readily available, they may continue using the previous version(s) consistent with the Instructions for Use, which require users to inspect the connecting tube prior to use. Transition to the current version as soon as possible by placing an order through Customer Service at 1-800-848-9024, option 2. When sufficient quantities of the current version (black) are obtained to support reprocessing needs, stop using the previous version and only use the current version going forward. Then arrange for return the previous version of the connector. Consignees were also requested to confirm receipt of the letter through the Olympus web portal and forward the notification to other users who may have the affected products if product were further distributed.
Details
- Recall number
- Z-1902-2026
- Initiated
- March 25, 2026
- Posted by FDA
- April 22, 2026
- Product code
- FEB – Accessories, Cleaning, For Endoscope
- Quantity
- 2,143 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Hong Kong, Japan, South Korea and Taiwan.
- Recalling firm
- Olympus Corporation of the Americas, Center Valley, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1898-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be…Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1899-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2111Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1900-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2112Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1901-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1903-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.XXX | 03/25/2026Open, Classified |
| Z-1904-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1905-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1906-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1907-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2119.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1908-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2138.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |
| Z-1909-2026 Class II | OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. | 03/25/2026Open, Classified |