Recall Z-1909-2024
Stryker Neurovascular · initiated April 23, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330
Codes / lots: Catalog/UDI-DI/Lot: M00313010/07613252186922/123279, 123364, 123404, 123985, 124214, 124216; M00313310/07613252186946/123494; M00316310/07613252187158/123851; M00316330/07613252187165/124156
Reason for recall
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
FDA-determined cause: Process design
Action taken
On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 7) Complete and return the response form via email to nvfieldactions@stryker.com
Details
- Recall number
- Z-1909-2024
- Initiated
- April 23, 2024
- Posted by FDA
- May 28, 2024
- Product code
- DQX – Wire, Guide, Catheter
- Quantity
- 2859
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland…
- Recalling firm
- Stryker Neurovascular, Fremont, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1902-2024 Class II | Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20…Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1903-2024 Class II | Trevo ProVue, Catalog: 90184Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1904-2024 Class II | Trevo Pro 14 Microcatheter, Catalog: 90231; Trevo Pro 18 Microcatheter, Catalog: 90238Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1905-2024 Class II | Trevo NXT ProVue 3 x 32, US, Catalog: 90312 Kits: TREVO NXT 3X32+TRAK21 2-PACK - CE…Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1906-2024 Class II | Trevo Trak 21 Microcatheter, Catalog: 90338Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1907-2024 Class II | FlowGate2 8F x 85cm, Catalog: 90485 FlowGate2 8F x 95cm, Catalog: 90495Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |
| Z-1908-2024 Class II | DAC 044 115cm, Catalog: 90760Neurovascular devices did not meet the minimum sample size required for endotoxin testing during… | 04/23/2024Open, Classified |