Recall Z-1918-2018

LivaNova USA · initiated March 13, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Inspire 6 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050713 The intended uses for the two elements that constitute the oxygenator/reservoir integrated devices are: INSPIRE 6M Hollow Fiber Oxygenator: The INSPIRE 6M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. The INSPIRE 6M is intended to be used for 6 hours or less. INSPIRE HVR (Hardshell Venous/Cardiotomy Reservoir): The INSPIRE HVR is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, de-foams and filters venous blood and suction blood. INSPIRE HVR can be used post-operatively for chest drainage. The INSPIRE HVR is intended to be used for 6 hours or less

Codes / lots: Lot: 1709190228, UDI: (01)08033178112352(17)200917(10)1709190228, Exp: 09/17/2020

Reason for recall

Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.

FDA-determined cause: Incorrect or no expiration date

Action taken

On March 13, 2018, notification letters were emailed. Customers were instructed to return response forms, which indicated the following: amount of unused product, and whether they would contact the firm to relabel effected products or if they preferred to be trained on the relabeling procedure. If product was further distributed, customers were instructed to send the notification letter to these additional customers within 2 business days. No action is required for product already used since it was consumed prior to the true expiration date. For further questions, please call (281) 228-7260.

Details

Recall number
Z-1918-2018
Initiated
March 13, 2018
Product code
DTZ – Oxygenator, Cardiopulmonary Bypass
Quantity
42
Distribution
US Distribution to the states of : NJ, CA, MN, MO, MA
Recalling firm
LivaNova USA, Arvada, CO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1919-2018 Class IIInspire 8 Integrated Phisio OXY w/ HVR Reservoir, Sterile EO, Catalog # 050714 The…Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique… 03/13/2018Terminated