Recall Z-1924-2013

Siemens Healthcare Diagnostics, Inc. · initiated July 1, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.

Codes / lots: TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14.

Reason for recall

Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.

FDA-determined cause: Process control

Action taken

Siemens sent an Urgent Medical Device Recall letter dated July 2013 to affected customers The letter identified the affected product, problem, potential risk, and actions to be taken by customers. Customers were requested to discuss the letter with thier laboratory director to determine if additional follow up is appropriate for patient tests conducted with the affected device lots. Additionally, customers were asked to discard any remaining inventory and to complete and return the field correction effectiveness check for replacement product. They were also asked to forward the letter to anyone wha had received this affected product. For questions contact your Siemens Technical Solutions Center or your local Siemens Technical support representative.

Details

Recall number
Z-1924-2013
Initiated
July 1, 2013
Posted by FDA
August 7, 2013
Terminated
August 26, 2014
Product code
MLM – Enzyme Immunoassay, Tracrolimus
Quantity
5423
Distribution
USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE

Official FDA recall record