Recall Z-1924-2025
Medtronic, Inc. · initiated April 29, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Azure S DR MRI SureScan, Product number W3DR01
Codes / lots: Serial numbers RNJ108800G, RNJ108808G, RNJ108810G UDI-DI 00763000956004 ***Updated 10/14/25*** Scope expanded to include implanted devices, serial numbers: RNJ108803G RNJ108804G RNJ108820G
Reason for recall
Identified devices may experience sudden battery power failure without alarm, power supply power-on-resets, and interruption of therapy. Potential patient harms include insufficient pacing support leading to cardiac arrest, syncope, or hospitalization; or system revision (explant/replacement) leading to infection, hematoma, or temporary impairment.
FDA-determined cause: Process change control
Action taken
Firm notified consignees of the recall on April 29, 2025 via letters titled "Urgent Medical Device Recall". Customers were instructed to identify the affected product and return to Medtronic. If you have any questions regarding this communication, please contact your Medtronic Field Representative ***Updated 10/14/25*** Firm expanded action to include three implanted devices. Consignees were notified on September 29, 2025. Consignees were instructed to utilize the CareLink remote monitoring system and inform a Medtronic Representative if a return of patient symptoms occurs or any unexpected device behaviors occur, including but not limited to: device resets or unexpected pacing diagnostics, with or without an alert. If the clinician assesses that prophylactic replacement of the device is necessary, standard warranty will apply. Medtronic will provide a replacement Azure S DR and a TYRXTM absorbable antibacterial envelope at no cost.
Details
- Recall number
- Z-1924-2025
- Initiated
- April 29, 2025
- Posted by FDA
- June 6, 2025
- Product code
- NVZ – Pulse Generator, Permanent, Implantable
- Quantity
- 6
- Distribution
- International distribution of the country of UK. No US distribution.
- Recalling firm
- Medtronic, Inc., Mounds View, MN