Recall Z-1931-2009

Abiomed, Inc. · initiated April 23, 2009

Terminated

Product

iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.

Codes / lots: Serial numbers: SC1000, SC1001, SC1003 - SC1010, SC1014 - SC1023, SC1025 - SC1070, SC1072 - SC1081, SC1083, SC1084, and SC1089

Reason for recall

iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.

FDA-determined cause: Device Design

Action taken

An "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.

Details

Recall number
Z-1931-2009
Initiated
April 23, 2009
Posted by FDA
September 16, 2009
Terminated
June 5, 2014
Product code
DSQ – Ventricular (Assist) Bypass
Quantity
79
Distribution
Domestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record