Recall Z-1943-2015

US Endoscopy Group Inc · initiated February 17, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.

Codes / lots: M/N: 00711140; Lot #'s: 1401267, 1404503, 1408953, 1413362, 1417521, 1420739, 1422288, 1422747, 1424014, and 1500443.

Reason for recall

The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation of the components during use.

FDA-determined cause: Under Investigation by firm

Action taken

US Endoscopy sent an Urgent Recall Notification Letter dated to their customers. The letter identified the affected product, problem and actions to be taken. US Endoscopy requests that their customers: 1. Remove all unused affected Velocity Biopsy Valves from your inventory. 2. Complete the attached Customer Response Card and return to the attention of Kia Pennington, via fax 1-440-639-4495 or email kpenning@usendoscopy.com. 3. Mark the shipping container with code: RMA xxxxxxxx QIF, and return the product to US Endoscopy via Fed Ex Account #293799547. For questions contact Kia Pennington in Customer Service at 1-800-769-8226.

Details

Recall number
Z-1943-2015
Initiated
February 17, 2015
Posted by FDA
July 1, 2015
Terminated
September 21, 2015
Product code
OCX – Endoscopic Irrigation/Suction System
Quantity
2,173 units
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, CO, IA, MA, OH, TX & WV and the countries of: Brazil, Costa Rica, Croatia, Czech Republic, Egypt, France, Germany, India, Indonesia, Italy, Latvia, Lithuania, Philippines, Poland, Romania, Russian Federation, Slovenia & South Africa.
Recalling firm
US Endoscopy Group Inc, Mentor, OH