Recall Z-1955-2021

Synaptive Medical Inc · initiated April 23, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Synaptive Trackable Suction Set Standard and Malleable

Codes / lots: Set and components codes: SYN-0657, SYN-0783 Consumables 3 Packs SYN-0663, SYN-0664, SYN-0665, SYN-0666, SYN-0667, SYN-0668, SYN-0759, SYN-0760, SYN-0761, SYN-0762, SYN-0763, SYN-0764, SYN-0765 and SYN-0766. Consumables Singles SYN- 0651, SYN-0652, SYN-0653, SYN-0654, SYN-0655, SYN-0656, SYN-0671,SYN-0672, SYN-0673, SYN-0674, SYN-0675, SYN-0676, SYN-0677 and SYN-0778. All lots and all serial numbers.

Reason for recall

Due to visible burrs/metal filings affixed along the suction tube inner perimeter.

FDA-determined cause: Vendor change control

Action taken

On April 23, 2021 the firm sent a letter to its consignees with the following instructions: 1. Quarantine and discontinue use of all Synaptive Trackable Suction devices. 2. Contact Synaptive Recall Support at 1.866.853.8001 or by email at RecallSupport@synaptivemedical.com. Customer service will schedule an on-site inspection of each device. If this is not possible, arrangements will be made to have all impacted units returned to Synaptive. 3. Once on-site inspection or return of your stock has been arranged, please complete and return the attached Acknowledgement and Receipt Form. All forms should be submitted by email to RecallSupport@synaptivemedical.com. Note: If you are in North America and your account is supported by a Synaptive Clinical Applications Specialist (CAS), then a company representative may have already visited your site and conducted the necessary inspections. If so, we kindly ask that you still complete the Acknowledgement and Receipt Form for our records. 4. Upon inspection, all units will be graded as pass or fail. Units that pass meet normal specifications and are safe for continued use. These devices will be returned and released back into circulation. Failing (i.e. defective) units will be removed immediately and replaced by Synaptive at no cost.

Details

Recall number
Z-1955-2021
Initiated
April 23, 2021
Terminated
January 12, 2024
Product code
GCX – Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Quantity
1291 units
Distribution
Domestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam
Recalling firm
Synaptive Medical Inc, Toronto

Official FDA recall record