Recall Z-1957-2013

SpineFrontier, Inc. · initiated May 9, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.

Codes / lots: Lot Number: 011836-0001, 011981-001

Reason for recall

The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.

FDA-determined cause: Device Design

Action taken

On May 9, 2012 an email was sent to consignees requesting that the product be returned for servicing and that a replacement part would be sent.

Details

Recall number
Z-1957-2013
Initiated
May 9, 2012
Posted by FDA
August 13, 2013
Terminated
January 10, 2014
Product code
HXX – Screwdriver
Quantity
9
Distribution
Distributed in DC and the states of MA, and OK.
Recalling firm
SpineFrontier, Inc., Beverly, MA

Official FDA recall record