Recall Z-1957-2016

Miami Fat Supply, Inc · initiated April 21, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

The Red Head 2 Reusable Fat Harvesting Canister, Product # RH-2, Non-Sterile. Intended for use in aspirating subcutaneous fatty tissue in patients desiring aesthetic body contouring and autologous fat collection.

Codes / lots: Lots 1165 and 1271.

Reason for recall

During an FDA inspection it was found that the products are marketed without a cleared 510k, nor is approved for multiple use.

FDA-determined cause: No Marketing Application

Action taken

Customers were sent an Urgent: Medical Device Recall/Correction letter, dated 04/19/2016, on 04/21/2016 that served as a notification of the Red Head 2 voluntary recall. Customers were informed that there were previous complaints of compromised product integrity and performance following sterilization. Lids are subject to fracture and fail after steam sterilization. Customers were advised to destroy used product and to ship unused product back to the manufacturer. Customers were to return an acknowledgement form to confirm their notification of the voluntary recall.

Details

Recall number
Z-1957-2016
Initiated
April 21, 2016
Terminated
December 14, 2016
Product code
MUU – System, Suction, Lipoplasty
Quantity
443 units
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, FL, GA, IL, MD, MI, MO, NM, NY, OH, TX, UT, and VA; and, the country of Dominican Republic.
Recalling firm
Miami Fat Supply, Inc, Groveland, FL

Official FDA recall record