Recall Z-1957-2025

Microbiologics Inc · initiated May 27, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC 17960; 0621K, Staphylococcus aureus derived from NCTC 12493

Codes / lots: 01065K: UDI-DI 30845357032868, lot 1065-25-51; 0621K: UDI-DI 30845357014666, lot 621-73-21

Reason for recall

Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.

FDA-determined cause: Mixed-up of materials/components

Action taken

On May 27, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed of the affected products containing the wrong microorganisms. Customers were instructed to review lab procedures to understand how this information affects their usage, and then to use or discard the product depending on the lab procedures and how the issue affects usage. Distributors of the affected product were instructed to contact end users who have received the affected items.

Details

Recall number
Z-1957-2025
Initiated
May 27, 2025
Posted by FDA
June 11, 2025
Product code
JTR – Kit, Quality Control For Culture Media
Quantity
3 total
Distribution
US Nationwide distribution in the states of California and Pennsylvania and the country of New Zealand.
Recalling firm
Microbiologics Inc, Saint Cloud, MN

Official FDA recall record