Recall Z-1960-2009

Stryker Spine · initiated June 14, 2006

Terminated

Product

Stryker Reflex Hybrid Variable Self-Tapping Screw. Non Sterile. Manufactured by Stryker Spine, SAS, Cestas, France. Distributed in USA by Howmedica Osteonics, Mahwah, NJ 07430.

Codes / lots: All Product with the following Catalog Number/Description: 48684012, Variable Angle Bone Screw, diameter 4.0 x 10 mm; 48684012, diameter 4.0 x 12 mm; 48684014, diameter 4.0 x 14 mm; 48684018, diameter 4.0 x 18 mm; 48684020, diameter 4.0 x 20 mm; 48684512, diameter 4.5 x 12 mm; 48684516, diameter 4.5 x 14 mm; 48684516, diameter 4.5 x 16 mm; 48684518, diameter 4.5 x 18 mm; 48684520, diamter 4.5 x 20 mm.

Reason for recall

Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.

FDA-determined cause: Device Design

Action taken

Stryker Spine issued an "Important Market Withdrawal" notice dated June 14, 2006. Consignees were advised to complete and return a Product Accountability Form via fax to 1-201-831-6069 prior to returning the affected product. Product return was to be performed during certain phase periods when upgrades were being shipped. For further questions, contact Stryker Spine at 1-201-831-5825.

Details

Recall number
Z-1960-2009
Initiated
June 14, 2006
Posted by FDA
August 20, 2009
Terminated
August 20, 2009
Product code
KWQ – Appliance, Fixation, Spinal Intervertebral Body
Quantity
108572 total self drilling and self tapping.
Distribution
Nationwide Distribution -- Including states of NC, OR, AZ, TN, CA, MA, MD, IL, GA, FL, NJ, NM, KY, LA, PA, MI, UT, MO, OK, KS, OH, MN, NE, IN, MT, TX, AL, CO and WI.
Recalling firm
Stryker Spine, Allendale, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1958-2009 Stryker Reflex Hybrid Level Anterior Cervical Plate; Non Sterile; Manufactured by…Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All… 06/14/2006Terminated
Z-1959-2009 Stryker Reflex Hybrid variable self-drilling screw. Non Sterile. Manufactured by Stryker…Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All… 06/14/2006Terminated