Recall Z-1978-2026
AVID Medical, Inc. · initiated March 13, 2026
Class I – dangerous or defective products that could cause serious health problems or death
Product
Halyard SAMMC ANGIOGRAPHY kit. Model Numbers: SAMM066-15.
Codes / lots: Model Number: SAMM066-15. UDI-DI (Kit Lot Numbers): 10809160388495 (1561685, 1575926, 1583048, 1584017, 1589652, 1595715, 1599634, 1605749, 1611810, 1621476, 1637222, 1637224, 1637225, 1643118, 1645143, 1645637, 1663929, 1663930, 1663931, 1663932).
Reason for recall
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
FDA-determined cause: Under Investigation by firm
Action taken
Owens and Minor/Avid Medical notified consignees on about 03/13/2026 and then sent an updated letter on about 03/18/2026 with additional kits added to the affected product list. The notice explained the issue was initiated by Medline Industries, risk associated with syringe disconnection and loose connection, and instructed consignees to identify, segregate and quarantine all affected product, and to apply warning labels to affected AVID kits to ensure all users are aware of the actions requested by Medline. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure. All usage guidelines and the Instructions for Use are provided in Medline s recall letter. Consignees were also instructed to complete and return the Recall Response Form as well as to notify customers if affected kits were further distributed to ensure all end-users are appropriately notified and maintain records of effectiveness.
Details
- Recall number
- Z-1978-2026
- Initiated
- March 13, 2026
- Posted by FDA
- May 1, 2026
- Product code
- OES – Cardiac Catheterization Kit
- Quantity
- 2,392 kits
- Distribution
- US distribution to FL, IL, MO, MS, NC, NE, TX.
- Recalling firm
- AVID Medical, Inc., Toano, VA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1970-2026 Class I | Halyard CARDIAC CATH PACK kit. Model Number: EAMC1000-05, WAFB208-02.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1971-2026 Class I | Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1972-2026 Class I | Halyard CARDIAC CATH TRAY SHANDS JAX kit. Model Number: SHND401-04, SHND401-05.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1973-2026 Class I | Halyard CATH LAB kit. Model Number: SACL75AM.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1974-2026 Class I | Halyard CATH LAB kit. Model Number: SACL75-01.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1975-2026 Class I | Halyard HEART CATH, SELF REGIONAL kit. Model Number: SELF131-05.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1976-2026 Class I | Halyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1977-2026 Class I | Halyard PERC TRAY kit. Model Numbers: SLPC34-01, SLPC34-02.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |
| Z-1979-2026 Class I | Halyard VA DETROIT CATH LAB PACK kit. Model Numbers: VMCD029.Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose… | 03/13/2026Open, Classified |