Recall Z-1981-2026

American Contract Systems Inc · initiated March 19, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ

Codes / lots: ANCA80AP UDI-DI 191072232168 Lot 42-8146111 ANCA80AQ UDI-DI 191072236678 Lots 42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911

Reason for recall

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

FDA-determined cause: Nonconforming Material/Component

Action taken

Firm notified distributors with an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" dated March 19, 2026, via email. Distributors were instructed to notify end-users. An updated letter with additional kits was sent to the customer on 3/20/26. Customers should quarantine all affected product. Customers may add warning labels to affected ACS kits to ensure all users are aware of prescribed actions that were identified in the Medline recall notice. The warning labels state, "Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure." Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, contact the Regulatory Affairs Manager, by email: gm-omra_recalls@owens-minor.com or at 419-455-2090.

Details

Recall number
Z-1981-2026
Initiated
March 19, 2026
Posted by FDA
May 6, 2026
Product code
LRO – General Surgery Tray
Quantity
7,311 kits
Distribution
US distribution to MN. No OUS distribution.
Recalling firm
American Contract Systems Inc, Tiffin, OH

Official FDA recall record