Recall Z-1994-2025

Barco N.V. · initiated May 9, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system

Codes / lots: REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726…

Reason for recall

Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures

FDA-determined cause: Nonconforming Material/Component

Action taken

On 5/9/2025, recall notices started being distributed to customers informing them of the following: 1) Assess & locate the non-installed affected units and contain them from further distribution. B) Assess & locate affected units that are installed at your customer side. Please make your customers aware of this notification. C) The impacted units will need to be sent back. Further information on the swaps will be provided in the coming days. D) Fill in and return the Acknowledgement Letter. If you don t have the Key Account Manager s contact details, or need other support, you can contact the firm 24/7 via their Support web page: https://www.barco.com/en/support

Details

Recall number
Z-1994-2025
Initiated
May 9, 2025
Posted by FDA
June 20, 2025
Product code
DXJ – Display, Cathode-Ray Tube, Medical
Quantity
221
Distribution
US Nationwide distribution in the states of TX, OH.
Recalling firm
Barco N.V., Kortrijk

Official FDA recall record