Recall Z-1997-2013

Beckman Coulter Inc. · initiated July 17, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.

Codes / lots: Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 19Sep2012 03/31/2014 M210079 04Oct2012 03/31/2014 M210169 27Oct2012 04/30/2014 M211100 10Nov2012 05/31/2014 M211078 01Dec2012 05/31/2014 M212122 12Dec2012 06/30/2014 M212203 03Jan2013 06/30/2014 M301324 16Jan2013 07/31/2014 M301418…

Reason for recall

Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.

FDA-determined cause: Under Investigation by firm

Action taken

Beckman Coulter sent an Urgent Product Correction letter dated July 9, 2013, to all affected customers. The letter was accompanied with a cover letter and response form, dated July 9, 2013. Customers were informed of the product under recall with product description, actions to take, contact information, the issue, impact, and resolution. Beckman Coulter conducted a risk assessment for this issue. Per the risk documentation, there is only a remote probability that use of the product will cause any adverse health consequences,and the product is not likely to lead to serious adverse health consequences. If you have any questions please call (714) 961-3777.

Details

Recall number
Z-1997-2013
Initiated
July 17, 2013
Posted by FDA
August 22, 2013
Terminated
October 24, 2014
Product code
KHE – Reagent, Occult Blood
Quantity
68,116 units
Distribution
Worldwide Distribution - USA and the country of Canada.
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record