Recall Z-2002-2020

ICU Medical, Inc. · initiated January 2, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Central Venous Catheters are indicated for the assessment of hemodynamic status through right atrial, right ventricular, and pulmonary artery and/or wedge pressure monitoring for patients. Model # 50456-01

Codes / lots: LOT Number/ UDI - 4350491/0 0840619 03011 7

Reason for recall

Inability for the guidewire to pass through the needles included with the catheter kits.

FDA-determined cause: Process control

Action taken

The firm has contracted Stericycle, Inc., Indianapolis, IN for customer notification, response tracking and effectiveness check. On 02 January 2020, Stericycle, Inc. sent URGENT: MEDICAL DEVICE RECALL notification of the market action to the firm's direct consignees of record.

Details

Recall number
Z-2002-2020
Initiated
January 2, 2020
Terminated
September 14, 2022
Product code
DQE – Catheter, Oximeter, Fiber-Optic
Quantity
15
Distribution
US Consignees including in the states of AL, CA, FL, GA, IL, ME, MI, MN, MS, NC, NH, NM, NY, OH, TX, UT, VA and WA OUS - Canada.
Recalling firm
ICU Medical, Inc., San Clemente, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1996-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated
Z-1997-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated
Z-1998-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated
Z-1999-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated
Z-2000-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated
Z-2001-2020 Class IICentral Venous Catheters are indicated for the assessment of hemodynamic status through…Inability for the guidewire to pass through the needles included with the catheter kits. 01/02/2020Terminated