Recall Z-2003-2026

Merit Medical Systems, Inc. · initiated April 2, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Philips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLY

Codes / lots: All Lot Numbers/UDI: 00884450816509

Reason for recall

catheter may experience resistance when being advanced over the guidewire

FDA-determined cause: Nonconforming Material/Component

Action taken

On April 17, 2026, Merit Medical Issued a Urgent Medical Device Recall notification via: FedEx. Initial notification was issued on April 2, 2026 via E-Mail and FedEx. Merit ask consignees to take the following actions: 1. Ensure that applicable personnel within your organization are made aware of this safety notice. 2. If the product has been further distributed to other facilities, institutions, or manufacturers, please immediately share this safety notice in accordance with your company policies. Please consider communicating the following to your customers: a. For continued safe use of the Philips Bridge" and Philips Bridge" Plus Occlusion Balloons, as outlined in the Prophylactic Directions for Use in the IFU, the following actions should be taken prior to the beginning of any procedure to confirm proper interaction between devices: i. After placement of the guidewire, advance the Balloon over the guidewire. This will confirm that the Balloon can freely be advanced over the entire length of the guidewire without resistance, thereby confirming that the guidewire is compatible with the Balloon. ii. Should resistance be felt during this step, the guidewire should be replaced with an alternate 0.035 stiff/super stiff wire, and Step 1 should be repeated prior to the beginning of the procedure. 3. Please circulate this information to all users of the Philips Bridge", Philips Bridge" Plus, and Bridge" Prep Kit device, or to any organization where the affected product may have been transferred, so that they are aware of the issue. Merit encourages you to instruct your customers to post the safety notice on or near the affected products.

Details

Recall number
Z-2003-2026
Initiated
April 2, 2026
Posted by FDA
April 29, 2026
Product code
OEQ – Angiography/Angioplasty Kit
Distribution
US Nationwide distribution in the state of CO.
Recalling firm
Merit Medical Systems, Inc., South Jordan, UT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2002-2026 Class IIPhilips Bridge Prep Kit REF: K12-09098B Sterile EO, Rx ONLYcatheter may experience resistance when being advanced over the guidewire 04/02/2026Open, Classified
Z-2004-2026 Class IIPhilips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLYcatheter may experience resistance when being advanced over the guidewire 04/02/2026Open, Classified