Recall Z-2024-2025
Limacorporate S.p.A · initiated May 19, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
PRIMA 2MM ECCENTRICAL ADAPTOR WITH SCREW REF: 1367.15.702, Sterile
Codes / lots: Lot number 2416658/UDI: 08033390201827
Reason for recall
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
FDA-determined cause: Process design
Action taken
On May 19, 2025, Enovis (Limacorporate S.p.A) issued a recall notification via e-mail. Enovis provided customers with an updated notification on May 28, 2025 to provide additional information. Enovis asked consignees to take the following steps: 1. Review your stock for the products and lot numbers for the items to be returned and quarantine immediately the affected devices. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. If you have further distributed the product, please provide this recall notification to those consignees; have them complete the acknowledgement form; and notify anyone that they have further distributed the affected product to. 3. Complete the attached acknowledgement form and contact Customer Service at 1-800-456-8696 or through your regional email address, to place a replacement request. You will be provided with an RMA number and shipping label to return the affected product. 4. Return all devices using the RMA number and shipping label provided by Customer Service. Devices must be sent back to: Lima U.S.A., Inc. | 2001 NE Green Oaks Blvd. Ste.100 | Arlington, Texas 76006 | USA within 15 days together with a signed hard copy of the following acknowledgement form.
Details
- Recall number
- Z-2024-2025
- Initiated
- May 19, 2025
- Posted by FDA
- June 25, 2025
- Product code
- PHX – Shoulder Prosthesis, Reverse Configuration
- Quantity
- 19 units
- Distribution
- Worldwide - US Nationwide distribution in the states of VA, FL, SC, MS, NY, TX and the countries of Italy, Poland, France.
- Recalling firm
- Limacorporate S.p.A, San Daniele Del Friuli