Recall Z-2027-2010

Lifescan Inc · initiated February 24, 2010

Terminated

Product

OneTouch SureStep and SureStep Pro Test strips, for use with the OneTouch SureStep Blood Glucose Meters, in 100 count and 50 count packaging distributed by LifeScan Inc., Milpitas, CA, Manufactured by LifeScan Puerto Rico Medical device for quantitative measurement of glucose in whole blood for use in home settings (SureStep).

Codes / lots: Part Number 02005201 (100 count), lot numbers 2969798, 2969251, 2982369, 2983467; Part number 01035904 (50 count), lot numbers 2969795 and 2982566; Part number 02096401 (50 count mail order pack), lot numbers 2969481 and 2998193

Reason for recall

Products may provide inaccurate results at readings over 400 mg/dL

FDA-determined cause: Other

Action taken

The firm, Lifescan, Inc., issued two "Urgent Medical Device Recall" letters dated February 24, 2010 to all customers for the OneTouch SureStep Test Strips and the SureStep Pro Test Strips. The letters describes the product, problem and action to be taken by the customers. The customers were instructed to look at your test strip vials for the recalled lots and check the strips or if you have test strips from the lot, call LifeScan Customer Service at 800-574-6139 between 5 a.m. to 7 p.m. Pacific Time, seven days a week or visit www.SureStep.com. Replacement product will be shipped to you immediately at no charge. The customers were instructed to continue to test your blood glucose, while wafting for replacement. Also, the customers are asked to continue use of the meter with precautions if results over 400 units are observed. For more information, please visit www.SureStep.com or call our Healthcare Professional Line at 888-455-5627.

Details

Recall number
Z-2027-2010
Initiated
February 24, 2010
Posted by FDA
July 17, 2010
Terminated
December 8, 2010
Product code
NBW – System, Test, Blood Glucose, Over The Counter
Quantity
1,014,350
Distribution
Worldwide distribution: USA and countries of Europe, Latin America and Asia.
Recalling firm
Lifescan Inc, Milpitas, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2028-2010 OneTouch SureStep Pro Test strips, for use with the OneTouch SureStep Pro Blood Glucose…Products may provide inaccurate results at readings over 400 mg/dL 02/24/2010Terminated