Recall Z-2028-2024

Remote Diagnostic Technologies Ltd. · initiated May 9, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Codes / lots: 00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior

Reason for recall

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

FDA-determined cause: Labeling design

Action taken

On 5/9/24, Philips distributed correction notices to customers and distributors asking them to do the following: 1) The User/Operator Manual was updated can be downloaded by searching "Tempus LS-Manual User-Operator Manual" at Philips.com/IFU 2) This notice needs to be shared with those who need to be aware within your organization and any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to ECR.Recall.Response@Philips.com. The following URL can also be used to complete/return the response form: https://philips.efrnfeedback.com/se/705E3ED80E91C689

Details

Recall number
Z-2028-2024
Initiated
May 9, 2024
Posted by FDA
June 6, 2024
Product code
LDD – Dc-Defibrillator, Low-Energy, (Including Paddles)
Quantity
1745
Distribution
US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.
Recalling firm
Remote Diagnostic Technologies Ltd., Farnborough

Official FDA recall record