Recall Z-2030-2021
The Binding Site Group, Ltd. · initiated May 20, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096
Codes / lots: Lot Numbers: 487719/ Exp: 31-Jul-2022, 480133/ Exp: 30-Sep-2022, Lot 457890/ Exp: 30-Sep-2022
Reason for recall
The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values.
FDA-determined cause: Under Investigation by firm
Action taken
Between the dates of 5/20-21/2021, The Binding Site (TBS) issued a "Urgent Medical device Correction" via email to affected consignees. In addition to informing consignees about the recall, the firm ask consignees to take the following actions: 1, TBS have completed reassignment of the controls; updated target values and assigned acceptance ranges are provided in Table 2. Users may continue to use remaining kits with the following revised assigned control values. 2. Please return you completed TSWS18 E-Back Form (attached) to technical.support@bindingsite.com or to your local Binding Site representative within 5 working days of receiving this notification. Transmission of this Important information: 3. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 4. Please transfer this notice to other organizations on which this action has an impact. 5. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action.
Details
- Recall number
- Z-2030-2021
- Initiated
- May 20, 2021
- Terminated
- February 4, 2025
- Product code
- DFH – Kappa, Antigen, Antiserum, Control
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, GA, FL, MA, MO, MI, WA, NY, PA, MN, AL, MD, OK, NE, IN, ID, OH, TX, HI, DC and the countries of Canada, Australia, Austria, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, Denmark, Estonia, France, Germany, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, United Kingdom…
- Recalling firm
- The Binding Site Group, Ltd., Birmingham